How to Manually Calibrate my FINCH 5332?
Titan Logix's older FINCH displays — the FINCH 5332E and FINCH 5332 IN-CAB — provide an LED numeric display of volume information, alarms, and system error codes from a TD80 or TD100 transmitter. Both displays can be manually calibrated using the steps below.
Applies To
FINCH 5332E and FINCH 5332 IN-CAB
FINCH 5332E
FINCH 5332 IN-CAB
Procedure
Manual Offset Calibration
Complete the following steps to manually calibrate your FINCH 5332E and/or FINCH 5332 IN-CAB display.
- Hold down an arrow button (it doesn't matter if it's the up or down arrow) on the front of the display, then turn on the power.
- Continue to hold the arrow button until you see the word CAL on the display.
- Once you see CAL on the display, release the arrow button.
- When the display shows the volume, press the up or down arrows on the display to adjust the display volume to match the measured amount of fluid.
- Turn the power off and back on again to complete the offset calibration.
Caution: There needs to be at least 6 inches of liquid in the tank, or the display will show 2Lo because of a low tank level.