You may come across the volume displayed on your FINCH suddenly starts jumping around and the values the FINCH is displaying do not make sense. This is an indication there is an issue with your TD System that needs to be addressed.
There are some things that you can go through yourself to try and diagnose this issue. Below are some of the things to look into further when you are faced with this issue:
-
Option 1: Probe Issue. Something might be stuck to the probe, your probe might be bent, or the shortening block might be loose at the bottom of tank.
-
Option 2: Connection between Transmitter and Probe. Your Transmitter might not be properly installed (connected) to the probe
- Option 3: Power supply issue. Check the power supply to the FINCH
- Option 4: If nothing above addressed this issue, there is an internal issue with your transmitter and needs to be repaired or replaced.
Need more help?
Titan Logix's Service Team provides 24-hour technical assistance for our customers.
Phone: 1-877-462-4085